STRYKER GMBH WIRE BOLT - MEDIUM HOFFMANN LRF FOR WIRE Ø1.5 TO 2.0MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 4933-1-002
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
The following event was reported by the stryker rep: "the hoffmann lrf frame was put on the patient on (b)(6) 2021 for an ankle fusion procedure to protect the internal hardware.Patient notified the surgeon on (b)(6) 2022 complaining of a loud 'pop' or a loose 'clicking sound.Frame was removed on the following day (b)(6) 2022.Every wire broke and nuts loosened up.The surgeon mentioned the patient was not walking on it throughout the few weeks it was on.".
 
Additional Manufacturer Narrative
This device is concomitant and did not contribute to the reported failure.Therefore, this complaint is closed without further investigation.If any additional information is provided indicating otherwise the record will be reopened and the investigation will be reworked.
 
Event or Problem Description
The following event was reported by the stryker rep: "the hoffmann lrf frame was put on the patient on (b)(6) 2021 for an ankle fusion procedure to protect the internal hardware.Patient notified the surgeon on (b)(6) 2022 complaining of a loud 'pop' or a loose 'clicking sound.Frame was removed on the following day (b)(6) 2022.Every wire broke and nuts loosened up.The surgeon mentioned the patient was not walking on it throughout the few weeks it was on.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WIRE BOLT - MEDIUM HOFFMANN LRF FOR WIRE Ø1.5 TO 2.0MM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13423873
Report Number0008031020-2022-00032
Device Sequence Number14441408
Product Code KTT
UDI-Device Identifier07613252507796
UDI-Public07613252507796
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K113327
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number4933-1-002
Device Catalogue Number49331002
Device Lot NumberG12007
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer02/16/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date03/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexMale
-
-