• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS SAFE-T-PRO PLUS; LANCET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS SAFE-T-PRO PLUS; LANCET Back to Search Results
Model Number SAFE-T-PRO PLUS
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer's lancets were requested for investigation and replacement product was sent to the customer.The occupation is patient/consumer.
 
Event or Problem Description
It was reported that a patient received an order of accu-chek safe-t-pro lancets and they were contaminated with mold.There was green and brown mold on the inside of the packaging and on the package insert.
 
Additional Manufacturer Narrative
One bag with 5 unused lancet devices were returned for investigation.The contamination with mold could be confirmed during the investigation.The affected lot number was packaged in (b)(6) 2021 and the batch record shows no record of any problems related to the device.The production process was reviewed and no abnormalities could be observed.The root cause of the contamination could not be identified.All bagged products are 100% visually checked.It is highly unlikely the issue occurred during the packaging process.The issue is consistent with contamination occurring either at customer site or at a third party.The investigation could not identify a product problem.The cause of the event could not be determined.
 
Additional Manufacturer Narrative
G1 contact office-manufacturing site was updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE-T-PRO PLUS
Common Device Name
LANCET
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
FACET TECHNOLOGIES LLC
112 town park drive suite 300
na
kennesaw GA 30144
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13424495
Report Number1823260-2022-00275
Device Sequence Number19831545
Product Code FMK
UDI-Device Identifier00365702702196
UDI-Public00365702702196
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/31/2025
Device Model NumberSAFE-T-PRO PLUS
Device Catalogue Number03450074001
Device Lot Number41821035
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer03/22/2022
10/16/2025
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date04/14/2022
11/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-