Model Number ORT100
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/25/2020
Event Type  malfunction  
Event Description
The customer service engineer observed a hydraulic fluid leak in one ptfe hose inside the ort100 table control panel in or2.There was no patient involvement.
 
Manufacturer Narrative
The customer service engineer observed a small hydraulic fluid leak in the ptfe hose in the control panel for the ort100 table.The leak did not impact the function of the table.The hose was replaced and the table was tested to verify proper function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT100 OPERATING ROOM TABLE
Type of Device
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer (Section G)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
kwaku amoah
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key13424905
MDR Text Key284888170
Report Number3010326005-2020-00015
Device Sequence Number1
Product Code FQO
UDI-Device Identifier00857534006332
UDI-Public00857534006332
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Biomedical Engineer
Remedial Action Repair
Type of Report Initial
Report Date 07/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORT100
Device Catalogue Number110470-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/25/2020
Initial Date FDA Received02/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-