| Model Number |
SKYRA |
| Medical Device Problem Codes |
Unintended Collision (1429); Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no device malfunction.A trained cleaning personnel failed to follow instructions and inadvertently brought a mr-unsafe trolley into the or suite while the magnet was deployed.The trolley was drawn into the magnet and caused damage to the magnet bore.The damaged magnet cover and collision detection ring were replaced and the mr was tested to verify proper function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, it was reported that a cleaning personnel was in the magnetic resonance suite while the magnet was deployed.An mr-unsafe trolley was drawn into the magnet, damaging the funnel cover and collision detection ring.There was no patient involvement and the user did not sustain any injury.
|
| |
|