Model Number COBAS 8000 CORE
Medical Device Problem Codes No Device Output (1435); Patient Data Problem (3197)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer was not performing daily "data backup and clear" because they needed to save 1 week of data in the core unit.The customer performed a data clean-up.They archived the results in data manager which resolved the issue.The investigation is ongoing.
 
Event or Problem Description
The initial reporter alleged a possible sample mismatch issue with a cobas 8000 core unit.The reporter stated that the patient sample ids were reused and the names on the patient result printouts from the core unit (cu) were incorrect.The instrument cu and result printout from cu are displaying a previous patient name.However, the results crossing to the laboratory information system (lis) were correct.There were no adverse events reported as the results crossed to the lis correctly.
 
Additional Manufacturer Narrative
No further issues were reported by the customer.The investigation determined the customer's actions (performance a data clean-up and archiving of the results in data manager) resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 CORE UNIT
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13425118
Report Number1823260-2022-00277
Device Sequence Number17141944
Product Code JJE
UDI-Device Identifier04015630928330
UDI-Public04015630928330
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K100853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCOBAS 8000 CORE
Device Catalogue Number05641446001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2022
Supplement Date Received by Manufacturer02/03/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-