Model Number 3100B
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A vyaire field service representative(fsr) evaluated the device onsite and verified that the unit fails with the delta p at 105.Calibration procedure was performed and found that the 73c peak to peak was low at around 50vp-p.The voltage was adjusted and all other parameters were verified.The patient circuit test passes and also passes the ventilator performance check with the delta p at 133.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
It was reported to vyaire medical that the 3100 b ventilator amplitude setting was only reading 104-105 cm.The customer confirmed that there was no patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Common Device Name
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL INC.
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key13425529
Report Number2021710-2022-15361
Device Sequence Number17474719
Product Code LSZ
UDI-Device Identifier10846446003543
UDI-Public(01)10846446003543(11)20020426
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P890057
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Initial Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2002
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1