| Model Number |
3100B |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A vyaire field service representative(fsr) evaluated the device onsite and verified that the unit fails with the delta p at 105.Calibration procedure was performed and found that the 73c peak to peak was low at around 50vp-p.The voltage was adjusted and all other parameters were verified.The patient circuit test passes and also passes the ventilator performance check with the delta p at 133.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the 3100 b ventilator amplitude setting was only reading 104-105 cm.The customer confirmed that there was no patient involvement associated with the reported event.
|
| |
|