EXACTECH, INC. LOGIC TIBIA IMPLANT PSC INSERT, SZ 4, 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number LOGIC TIBIA IMPLANT PSC INSERT, SZ 4, 15MM
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, patient presented with instability and a poly exchange was performed where the surgeon went to a thicker insert.The tibia and femur were well fixed.Images were received.No additional information available.
 
Additional Manufacturer Narrative
Concomitant medical products: logic femoral ps cem left sz 4 (cat# 02-010-01-0240 / serial# (b)(4)); lgc tibial fit tray cem sz 4f / 4t (cat# 02-012-45-4040 / serial# (b)(4)); three peg patella 35mm (cat# 200-02-35 / serial# (b)(4)); collar fixation pin 2pk 40mm, quick release (cat# 201-78-82 / serial# (b)(4)); 3" drill bit, mod.Hex 2 pack (cat# 201-78-89 / serial# (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Upon review of all available information, there is no evidence to suggest that a design or manufacturing issue caused or contributed to this event.The reported revision was likely the result of underlying patient factors, which led to instability of the joint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA IMPLANT PSC INSERT, SZ 4, 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key13426044
Report Number1038671-2022-00142
Device Sequence Number14451199
Product Code JWH
UDI-Device Identifier10885862173669
UDI-Public10885862173669
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K110547
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/08/2022
Device Model NumberLOGIC TIBIA IMPLANT PSC INSERT, SZ 4, 15MM
Device Catalogue Number02-012-44-4015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/10/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-