| Catalog Number |
MB3355 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the inferior endplate detached from the cartridge, causing the mobi-c implant to rotate into the improper placement.The surgeon removed and replaced the implant without a delay or patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information in h4 and h6: component, investigation type, findings, and conclusions.Device evaluation: product was not returned and photos were not provided, so a device evaluation could not be performed.Root cause: root cause was unable to be determined.This event could possibly be attributed to off-axis forces applied on the implant.It could also be attributed to over-turning the holder knob during assembly or insertion.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that the inferior endplate detached from the cartridge, causing the mobi-c implant to rotate into the improper placement.The surgeon removed and replaced the implant without a delay or patient impacts.
|
| |
|