Model Number M00545150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 49 was intended to be used in the common bile duct in an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During unpacking outside the patient, it was noticed that the cutting wire of the autotome rx 49 was broken.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6 (device codes): medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned autotome rx 49 was analyzed, and a visual evaluation noted that the guidewire introducer was broken.The device was observed under magnification and the cutting wire was in good condition.A functional evaluation was performed by bowing the device outside and inside the scope and the device bowed and returned from bowed position as intended.No other problems with the device were noted.The reported event of cutting wire break was not confirmed.Upon analysis, the cutting wire was in good condition.Additionally, it was found that the guidewire introducer was broken, which could have been caused by manipulation factors, such as user technique.The reported problem was noticed upon unpacking, so most likely, as part of the device manipulation during preparation, an excess of force was applied in order to get the device out of the package damaging the guidewire introducer.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 49 was intended to be used in the common bile duct in an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During unpacking outside the patient, it was noticed that the cutting wire of the autotome rx 49 was broken.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOTOME RX 49
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13427431
Report Number3005099803-2022-00395
Device Sequence Number5379668
Product Code KNS
UDI-Device Identifier08714729415138
UDI-Public08714729415138
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/06/2024
Device Model NumberM00545150
Device Catalogue Number4515
Device Lot Number0026590571
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/14/2022
Supplement Date Received by Manufacturer02/28/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
Patient Weight68 KG
-
-