| Model Number |
0684-00-0296-02 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during insertion of the intra-aortic balloon (iab), the balloon would not progress through sheath and artery even after several attempts.A new balloon was opened and progressed without problem.There was no reported injury or adverse event to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane loosely folded and traces of blood on the exterior.The sheath was not returned for evaluation.A laboratory insertion test was unable to be performed due to the membrane being unfurled.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.A leak may impact the ability to maintain vacuum.We are unable to confirm the reported difficulty during insertion because of the returned condition of the catheter and we are unable to mimic the clinical setting.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, d10, g4, h3, h6, h10.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
|
| |
|