MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED Back to Search Results
Model Number 123620A
Medical Device Problem Code Material Too Rigid or Stiff (1544)
Health Effect - Clinical Codes Cramp(s) /Muscle Spasm(s) (4521); Skin Inflammation/ Irritation (4545)
Date of Event 01/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the patient did not like the foley catheters, and they were not working for the patient.Per follow up information received via phone on 12jan2022, it was stated that the patient used a clear, hard, stiff catheter which caused irritation and spasms.The patient went to the doctor and was given a prescription medicine for the spasms and the patient switched back to the catheters that the patient usually uses.The patient did not know the name of the prescribed medicine.The patient¿s normal catheter was reddish brown in color.The patient had no further irritation issues since switching back to the usual catheter, but the patient was experiencing leakage.
 
Event or Problem Description
It was reported that the patient did not like the foley catheters, and they were not working for the patient.Per follow up information received via phone on (b)(6)2022, it was stated that the patient used a clear, hard, stiff catheter which caused irritation and spasms.The patient went to the doctor and was given a prescription medicine for the spasms and the patient switched back to the catheters that the patient usually uses.The patient did not know the name of the prescribed medicine.The patient¿s normal catheter was reddish brown in color.The patient had no further irritation issues since switching back to the usual catheter, but the patient was experiencing leakage.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.It is unknown whether the device had met relevant specifications.No sample was returned for evaluation.Potential root cause for this failure mode could be catheter shaft gauge is out of specification.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon to burst.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer syringe.Do not use needle.Visually inspect the product for any imperfections or surface deterioration prior to use.Do not use if package is opened or damaged.Recommended inflation capacities 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water do not exceed recommended capacities.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® FOLEY CATHETER SILICONE COATED
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13429008
Report Number1018233-2022-00336
Device Sequence Number11698517
Product Code KOD
UDI-Device Identifier00801741095214
UDI-Public(01)00801741095214
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number123620A
Device Catalogue Number123620A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer02/24/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date03/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-