The following information was reported to gore: on (b)(6) 2010, this patient underwent an emergency endovascular treatment for the ruptured abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.Reportedly, the aneurysm size was about 9 cm.On an unknown date, postoperative follow-up, a tendency of migration of the left leg/ipsilateral leg was confirmed.The amount of the migration was unknown.On (b)(6) 2021, follow-up ct scan showed migration of the left leg into the aneurysm sac, resulting in a distal type i endoleak.On (b)(6) 2022, a reintervention was performed.An additional stent graft was implanted into the left distal side.It was reported that the endoleak was fixed.The patient tolerated the procedure.It was reported that the size of the distal left leg at the index procedure may have been undersized, however, the treatment length for the index procedure was unknown.Additionally, no comment was received from the physician regarding the cause of migration.
|