Model Number B2060-040
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2021, a percutaneous intervention was performed on the superficial femoral artery (sfa) and popliteal, moderately calcified lesions.An armada 35 balloon catheter advanced to the sfa without issues.On the 1st inflation at 6 atmospheres (atms), the balloon burst.An armada 14 advanced to the popliteal without issues.Also, on the 1st inflation at 6 atms, the balloon burst.Both devices were removed without issue and another device was used in replacement.There was no adverse patient effect and no clinically significant delay.No additional information was provided regarding this issue.
 
Additional Manufacturer Narrative
The device was received.The investigation is not yet complete.A follow up report will be submitted with all additional relevant information.The other armada balloon catheter referenced is filed under a separate medwatch report number.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other complaints reported from this lot.Based on the information provided, the reported material rupture appears to be due to circumstances of the procedure.It is likely that the balloon rupture occurred due to interaction with lesion calcification.There is no indication of a product quality issue with respect to the design, manufacture, or labeling.D9, h3: device not available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13431600
Report Number2024168-2022-01154
Device Sequence Number4492742
Product Code LIT
UDI-Device Identifier08717648155260
UDI-Public08717648155260
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberB2060-040
Device Catalogue NumberB2060-040
Device Lot Number10907G1
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/13/2022
Supplement Date Received by Manufacturer02/28/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-