• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MIDAS REX® LEGEND ATTACHMENT; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MIDAS REX® LEGEND ATTACHMENT; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number AF02R
Medical Device Problem Codes Break (1069); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.Evaluation determined that attachment was damaged.The likely cause was identified as leg is worn and foot is chipped.The user manual contains the following warnings ¿do not use a legend attachment if any part of the attachment appears to be bent, loose, missing or damaged.Do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff." in addition, ¿insert dissecting tool into motor collet with a slight rotational motion.A tactile and audible click is observed indicating that the dissecting tool is fully seated.If damage is found, the device should not be used.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with the report of broken.It was reported that there was no patient involvement.Repair escalated to a product event due to evaluation finding of chipped foot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX® LEGEND ATTACHMENT
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
glen belmer
7000 central avenue ne
ft. worth, TX 76137-3291
6122713209
MDR Report Key13435707
Report Number1625507-2022-00015
Device Sequence Number4327770
Product Code HBB
UDI-Device Identifier00613994469168
UDI-Public00613994469168
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAF02R
Device Catalogue NumberAF02R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2022
Initial Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-