This device is classified as import for export, therefore 510k is not applicable.Model vnl11-j10 is available in the usa with a 510k number k172156.We checked the returned unit and confirmed teethe cfb flexible tube hardened.Based on the result, we concluded that it was caused due to overused.In addition, we confirmed that the insertion flexible tube (ift) crushed, the remote control buttons perforated, the remote control buttons leaky, and the u/d lock lever cracked ; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|