HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE Back to Search Results
Model Number VNL-1190STK
Device Problem Collapse (1099)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2022
Event Type  malfunction  
Event Description
The time of event is unknown.There was no report of patient harm.Insertion flexible tube (ift) hardened.
 
Manufacturer Narrative
This device is classified as import for export, therefore 510k is not applicable.Model vnl11-j10 is available in the usa with a 510k number k172156.We checked the returned unit and confirmed teethe cfb flexible tube hardened.Based on the result, we concluded that it was caused due to overused.In addition, we confirmed that the insertion flexible tube (ift) crushed, the remote control buttons perforated, the remote control buttons leaky, and the u/d lock lever cracked ; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key13435784
MDR Text Key295873050
Report Number9610877-2022-50421
Device Sequence Number1
Product Code EOB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVNL-1190STK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2022
Initial Date Manufacturer Received 01/20/2022
Initial Date FDA Received02/03/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-