C.R. BARD, INC. (BASD) -3006260740 LEONARD CV CATHETER, DUAL-LUMEN,10F; CHRONIC CATHETERS
|
Back to Search Results |
|
| Model Number |
0600340 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a chronic catheter placement procedure, the device was allegedly broken.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported fracture issue as no objective evidence was provided for review.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 01/2025).
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one leonard 10f d/l catheter and two luer caps were returned for evaluation.Gross, microscopic visual, tactile evaluation and functional testing were performed.The investigation is confirmed for the reported catheter fracture issue as a partial circumferential break was noted on the strengthening sheath approximately 3.8cm from the bifurcation and the edges of the break appeared to be uneven and surface appeared to be granular.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiry date: 01/2025).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a chronic catheter placement procedure, the device was allegedly broken.There was no reported patient injury.
|
| |
|