Model Number 393-090 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2021 |
Event Type
malfunction
|
Event Description
|
The customer noticed pco2 & ph deviation has occurred on (b)(6) 2021 as like following.18:06 ph 7.1, pco2 64.8 >ph lower & pco2 higher than dr.Exepected.18:22 ph 7.1, pco2 62.7 >ph lower & pco2 higher than dr.Exepected.18:30 ph 7.4, pco2 37.6 >ph & pco2 almost same as dr.Exepected.Fortunately this customer have not treat the patient.
|
|
Manufacturer Narrative
|
The prelimenary radiometer investigation has shown that several parameters deviate in the first 2 blood samples (18:06 and 18:22): ph, pco2, po2, ca, and na.No root cause established yet, the investigation is ongoing.
|
|
Manufacturer Narrative
|
The radiometer investigation has shown, that the abl90 flex analyzer worked as intented and the problem is caused by pre-analytical handling.The deviating measurements are most likely a consequence of the samples being collected using edta or citrate as anticoagulant and without proper protection from atmospheric air.The sample deviates on multiple parameters consistent with the use of edta or citrate as anticoagulant and the use of a sample vessel without proper protection from atmospheric air.
|
|
Search Alerts/Recalls
|
|