• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC SYMPHONY OCT SYSTEM DRILL GUIDE MODULAR HANDLE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC SYMPHONY OCT SYSTEM DRILL GUIDE MODULAR HANDLE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 2020-00-119
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter is synthes sales representative the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2022, after case, while associate was disassembling instruments for cleaning, the symphony drill guide handle disconnected from button and associated internal shaft.Associate noted the problem which could not be remedied and set was sent for wash and decontamination.There was no patient involvement.Procedure outcome is unknown.This report is for one (1) symphony oct system drill guide modular handle this is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.This complaint is confirmed as the device was not able to assemble as intended.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a review of the receiving inspection (ri) for drill guide modular handle was conducted identifying that lot number km894607 was released in one batch.Batch1: lot qty of units were released on 06 oct 2020 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMPHONY OCT SYSTEM DRILL GUIDE MODULAR HANDLE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
BRIDGEWATER DISTRIBUTION
50 scotland boulevard
bridgewater MA 02324
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13440089
Report Number1526439-2022-00175
Device Sequence Number4492437
Product Code LXH
UDI-Device Identifier10705034530413
UDI-Public(01)10705034530413
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2020-00-119
Device Catalogue Number202000119
Device Lot NumberKM894607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer02/09/2022
Initial Report FDA Received Date02/03/2022
Supplement Report FDA Received Date03/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/06/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN DRILL GUIDE BODY; UNKNOWN DRILL GUIDE SLEEVE
Patient SexUnknown
-
-