Model Number M00522600
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a resolution clip device was used in the stomach during an endoscopic submucosal dissection (esd) procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue, but could not be released from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Event or Problem Description
It was reported to boston scientific corporation that a resolution clip device was used in the stomach during an endoscopic submucosal dissection (esd) procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue, but could not be released from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: medical device problem code a15 captures the reportable event of clip unable to release from catheter.Block h10: investigation results the returned resolution clip was analyzed, and a visual evaluation noted that the clip assembly was returned attached to the catheter.The device was returned without its over sheath.Microscopic examination was performed, and it was observed that the control wire was disconnected from the handle spool.Functional evaluation, and it was found that between the handle and the clip assembly had lose connection.No other problems with the device were noted.The reported event was not confirmed.Device analysis found the control wire was disconnected from the handle and cause the handle had lose connection from the catheter.This action also cause the physician was not able to perform the first activation preventing the clip arms to close.Additionally, the control wire was inspected under microscope and it was observed that it looks like if it was not correctly inserted into the barbed crimp.This unit was manufactured on 09jun2021 before additional line controls were implemented which includes an update in the method in which the product builders insert the control wire in the cross pin in order to avoid the control wire to detach from the cross pin.All issues have been addressed.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION CLIP
Common Device Name
CLIP, HEMOSTATIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13440391
Report Number3005099803-2022-00482
Device Sequence Number4492443
Product Code MCH
UDI-Device Identifier08714729504764
UDI-Public08714729504764
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K040148
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/24/2024
Device Model NumberM00522600
Device Catalogue Number2260
Device Lot Number0027465380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer02/07/2022
Initial Report FDA Received Date02/03/2022
Supplement Report FDA Received Date03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
Patient Weight65 KG
-
-