| Model Number |
720170 |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer stated the patient had a button placed on (b)(6) 2021.While at home, the patient's peg button had a balloon failure and leaked following a tussive episode with vomiting.They went to the er where the new kangaroo 12fr was repositioned.It was filled with 2.5ml, reflux of gastric material occurred and the probe was functioning.The customer clarified there was no additional surgery, the device was replaced in the er.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary.The device history record (dhr) review could not be performed because the reported lot number is not a valid lot number.One decontaminated nutriport without its original package or lot number was received at the manufacturing plant for evaluation.It was visually inspected according to procedure.The balloon was filled with sterile water, and a leak was observed under the balloon.The reported condition was confirmed.A corrective and preventative action (capa) was opened to determine the root cause and corrective actions.This complaint will be used for tracking and trending purposes.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 lot number corrected & nbsp; from lot 20022114564 to lot 2107429764 per information received from the customer.
|
| |
|