Model Number 788628
Medical Device Problem Codes Material Fragmentation (1261); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, the catheter was replaced in the patient.The external package of the new product was checked prior to use and found intact without any abnormalities.However once after the external package was opened, the pigtail coil at the tip of the ureteral stent was found to be broken.The user opened a new kit of stent to complete the procedure.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Additional Manufacturer Narrative
The reported event was confirmed, however the cause was unknown.No physical sample was returned, however, photo samples were submitted.The position of the stent pictured appears to be in a bent position; the sample does not appear to be broken by the position of the stent pictured.A potential root cause for this event could be, "inappropriate package design".As no patient involvement was reported the device was not used on a patient, though it is intended for treatment purposes.As the device was not returned for evaluation, it was unknown if it had met all relevant specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that on (b)(6) 2022, the catheter was replaced in the patient.The external package of the new product was checked prior to use and found intact without any abnormalities.However once after the external package was opened, the pigtail coil at the tip of the ureteral stent was found to be broken.The user opened a new kit of stent to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13441221
Report Number1018233-2022-00356
Device Sequence Number19396040
Product Code FAD
UDI-Device Identifier00801741015779
UDI-Public(01)00801741015779
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K043193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/14/2025
Device Model Number788628
Device Catalogue Number788628
Device Lot NumberNGFP3233
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date02/03/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-