| Model Number |
HFD100 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
Skin Tears (2516)
|
| Date of Event |
11/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The torque screw and skull clamp table adapter were returned for evaluation.The torque screw was found to be functioning within specifications.The table adapter was found to be seized and full clamping could not be achieved.The adapter was unable to be disassembled for further inspection, but the seizing was most likely caused by internal deformation in the internal part, potentially by application of excess force or cross-threading.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
|
| |
|
Event or Problem Description
|
|
During a patient procedure on (b)(6) 2019, two neurosurgeons noticed the loose fit of the hfd100 skull clamp assembly.The patient's head moved during drilling however, the patient did not sustain any injuries other than skin damage.
|
| |
|