• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1.5MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1.5MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035421520
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2021
Type of Reportable Event Malfunction
Event or Problem Description
Analysis of the returned device found that the coil delivery wire broken/fractured during use.There were no clinical consequences to the patient reported as a result of this event.
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.Visual/microscopic inspection: the delivery wire was found to be broken/ fractured 11 cm from the proximal end.The delivery wire broke off completely during the analysis.The coil delivery wire was found to be kinked/bent.The main coil was found to be stretched as per photo.The suture was found to be intact.Functional inspection: main coil failed/unable to detach functional test: unable to perform the test as the delivery wire was found to be broken/fractured and broke off completely during analysis.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was not confirmed during device analysis however, the analysis results are consistent with the reported event.The device did not meet specifications when received for complaint investigation based on functional and visual inspection.Additional information provided by the customer indicated that there was no damage noted to the packaging prior to opening the packaging.The device was prepared for use as per the dfu and continuous flush was set up and maintained throughout the clinical procedure.It was reported that multiple attempts were made to detach the coil.The device was returned for analysis and the coil delivery wire was seen to be kinked and broken.The main coil was found stretched.The as analyzed codes 'coil delivery wire broken/fractured during use', 'coil delivery wire kinked/bent' and 'main coil stretched' will be assigned procedural factors as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.An assignable cause of 'not confirmed' was assigned to the aa reported defect 'main coil failed/unable to detach' as the issue included a returned product review (visual, physical, and/or performance testing) which showed no evidence of either the alleged issue(s) or any defect which could have contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1.5MM X 2CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13442067
Report Number3008881809-2022-00048
Device Sequence Number19396044
Product Code HCG
UDI-Device Identifier04546540697981
UDI-Public04546540697981
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2023
Device Model NumberM0035421520
Device Catalogue NumberM0035421520
Device Lot Number22435067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2022
Initial Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SL-10 MICROCATHETER (STRYKER); XT-17 MICROCATHETER (STRYKER)
Patient Age53 YR
Patient SexMale
Patient Weight68 KG
Patient RaceAsian
-
-