Catalog Number 532.101
Device Problems Noise, Audible (3273); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis. If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate. Additional narrative: device evaluation: the actual device was returned for evaluation. During repair, an evaluation was performed and it was determined that the trigger being sticky, identified during service and evaluation were confirmed. The assignable root cause was determined to be due to component failure from normal wear. Udi: (b)(4).
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was determined that the upper and lower triggers of the small battery drive device were stuck and sticky, and the device made a strange and an unexpected noise. It was further determined that the device failed pretest for check for sticky triggers. It was noted in the service order that the device made an abnormal noise. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of the event was unknown. However, it was reported that the event occurred in 2021. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOLIBRI II HANDPIECE
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ 4436
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6107428552
MDR Report Key13442833
MDR Text Key290381917
Report Number8030965-2022-00677
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 02/03/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number532.101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Date Manufacturer Received01/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-