SYNTHES GMBH COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.101 |
Device Problems
Noise, Audible (3273); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the trigger being sticky, identified during service and evaluation were confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was determined that the upper and lower triggers of the small battery drive device were stuck and sticky, and the device made a strange and an unexpected noise.It was further determined that the device failed pretest for check for sticky triggers.It was noted in the service order that the device made an abnormal noise.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|