• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83779
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.Visual inspection revealed that a valve and the pusher wire were returned for this complaint.The pusher wire was inspected, and it was damaged (kinked and stretched).Also, the interlocking arm was detached.No more damages were found.Microscopic inspection of the pusher wire revealed that the proximal end has a smooth surface.The interlocking arm was detached.
 
Event or Problem Description
Reportable based on device analysis completed on 10jan2022.It was reported that the delivery system could not advance the coil.The target lesion was located in the splenic artery.A 5mm x 15cm interlock was selected for use.During the procedure, it was noted that the delivery system could not advance the coil.The procedure was completed with another of same device.No patient complications were reported and patient was stable post procedure.However, device analysis revealed that the interlocking arm was detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13442885
Report Number2134265-2022-00892
Device Sequence Number2045687
Product Code KRD
UDI-Device Identifier08714729764984
UDI-Public08714729764984
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/08/2023
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0025336417
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/10/2022
Initial Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
Patient Weight67 KG
-
-