| Model Number |
SNGCB |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date of event: unknown; captured as awareness date.Medical device: batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an unknown procedure, the blade came off when the device was connected to the handpiece.No pieces fell into the patient.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/25/2022 investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.The device was received with the curve blade out off the shaft sheath.The event reported is confirm.No functional test could be perform due to the condition of the device returned, the handles was disassembled to inspect the internal components and was noted that the insulated retention pin was missing and not retuned.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery quality system.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances related to the reported complaint condition were identified.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.
|
| |
|
Search Alerts/Recalls
|