• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: XTANT MEDICAL HOLDINGS, INC. SILEX SACROILIAC JOINT FUSION SYSTEM; SACROILIAC JOINT FIXATION / SACROILIAC JOINT FUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

XTANT MEDICAL HOLDINGS, INC. SILEX SACROILIAC JOINT FUSION SYSTEM; SACROILIAC JOINT FIXATION / SACROILIAC JOINT FUSION Back to Search Results
Model Number X079-0090
Medical Device Problem Code Solder Joint Fracture (2324)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A visual assessment of the returned complaint guide handle assembly showed an instrument with repeated use, as identified by worn laser markings and surface scratches.There were deep scratches and impact marks identified on the outer ring as well as on the nut.The distal end of the handgrip was misshaped with extensive scratches and compression marks identified.A weld on the instrument base was fractured at the same location of the compression marks identified on the handgrip.The separated weld was misshaped in the same manner and location as the damage identified on the handgrip.A functionality assessment was not performed due to the damaged condition of the complaint instrument, which was removed from distributable inventory.A dhr review was performed for the guide handle assembly complaint lot and there were no manufacturing anomalies identified.The device met all required specifications prior to being released to distributable inventory.This device lot has been available for distribution since (b)(6) 2016.It may be possible for the handle to separate from the body of the instrument if excessive force was placed on the handle.It is unknown why excessive force was placed on the instrument handle.The damage identified on the handle suggests a plier-type instrument was used to grasp the instrument, which compressed a weld and contributed to the instrument malfunction.There have not been any other complaints of similar nature in the past 12 months.The manufacturer will continue to monitor this instrument for complaints from the field.
 
Event or Problem Description
The manufacturer received notification on (b)(6) 2022 of an instrument malfunction that occurred during a surgical procedure.It was reported that the handle of a guide handle assembly fractured from the body of the instrument.There were no known patient complications associated with this complaint and the procedure was successfully completed.The complaint instrument was returned to the manufacturer on (b)(6) 2022 for complaint assessment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILEX SACROILIAC JOINT FUSION SYSTEM
Common Device Name
SACROILIAC JOINT FIXATION / SACROILIAC JOINT FUSION
Manufacturer (Section D)
XTANT MEDICAL HOLDINGS, INC.
664 cruiser lane
belgrade MT 59714
Manufacturer (Section G)
XTANT MEDICAL HOLDINGS, INC.
664 cruiser lane
belgrade MT 59714
Manufacturer Contact
rebecca lennemann
664 cruiser lane
belgrade, MT 59714
MDR Report Key13443673
Report Number3005031160-2022-00012
Device Sequence Number11838442
Product Code OUR
UDI-Device IdentifierM697X07900901
UDI-PublicM697X07900901
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140079
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberX079-0090
Device Lot Number056994
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2022
Initial Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-