| Model Number |
N/A |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 00598004702, stemmed tibial component precoat size 6 for cemented use only use of this tibial component with lcck articulating surfaces requires using a stem exten, 64900483.Report source: foreign country: (b)(6).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the insert would not lock and fit properly when installed during the procedure following the proper surgical technique and tools.There was no surgical delay.No known impact or consequence to the patient.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 00598004702, stemmed tibial component precoat size 6 for cemented use only use of this tibial component with lcck articulating surfaces requires using a stem exten, 64900483.Report source: foreign country: (b)(6).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the insert would not lock and fit properly when installed during the procedure following the proper surgical technique and tools.There was no surgical delay.No known impact or consequence to the patient.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual evaluation of the returned implant shows signs of use.Dovetail feature was compressed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.Damage to the articular surface can occur if the articular surface is not correctly placed and oriented before pushing it using the inserter instrument.Detailed instructions for inserting the articular surface are provided in the surgical technique.The root cause is attributed to user error.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|