No product was received as the fractured portion of the device currently remains in situ and the remainder was not returned.Radiograph's were provided but could not confirm the fracture prior to attempted removal.Photographs provided of the fractured screw suggest a high reverse flexural load fracture.The root cause of the screw fracture is unknown but review of the information provided suggests excessive off angled force during removal as a probable cause.No additional investigation required.Labeling review ".Potential adverse events and complications - with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative nd postoperative complications that may result in the need for additional surgeries.Potential risks identified with the use of this system, which may require additional surgery, include: ending, fracture or loosening of implant component(s)¿" ".Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants.Metallic internal fixation devices cannot withstand the activity levels and/or loads equal to those placed on normal, healthy bone.These devices are not designed to withstand the unsupported stress of full weight or load bearing alone.Caution must be taken due to potential patient sensitivity to materials.Do not implant in patients with known or suspected sensitivity to the aforementioned materials." ".These devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient¿s activity level will dictate the longevity of the implant.".
|