NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 13015540
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product was received as the fractured portion of the device currently remains in situ and the remainder was not returned.Radiograph's were provided but could not confirm the fracture prior to attempted removal.Photographs provided of the fractured screw suggest a high reverse flexural load fracture.The root cause of the screw fracture is unknown but review of the information provided suggests excessive off angled force during removal as a probable cause.No additional investigation required.Labeling review ".Potential adverse events and complications - with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative nd postoperative complications that may result in the need for additional surgeries.Potential risks identified with the use of this system, which may require additional surgery, include: ending, fracture or loosening of implant component(s)¿" ".Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants.Metallic internal fixation devices cannot withstand the activity levels and/or loads equal to those placed on normal, healthy bone.These devices are not designed to withstand the unsupported stress of full weight or load bearing alone.Caution must be taken due to potential patient sensitivity to materials.Do not implant in patients with known or suspected sensitivity to the aforementioned materials." ".These devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient¿s activity level will dictate the longevity of the implant.".
 
Event or Problem Description
On (b)(6) 2022 a patient underwent a revision procedure to extend a pre existing construct at l3-l5 to t9-pelvis due to a continued decline in patient pathology.During the hardware removal portion of the revision surgery it was discovered that the right screw at l5 had fractured.As a consequence the fractured portion was left in patient.The remainder of the revision occurred without issue and the patient is currently recovering well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
san diego, CA 92121
MDR Report Key13444144
Report Number2031966-2022-00026
Device Sequence Number4492559
Product Code NKB
UDI-Device Identifier00887517536846
UDI-Public887517536846
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K143684
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number13015540
Device Lot NumberJP5257
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2022
Initial Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/27/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-