| Model Number |
CD2257-40 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Interrogation of the device revealed the device was at elective replacement indicator (eri) when received.There¿s no bpa detection.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Based on this information, there was no evidence of premature depletion.
|
| |
|