• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SEASPINE ORTHOPEDICS CORPORATION MARINER PEDICLE SCREW SYSTEM; SET SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SEASPINE ORTHOPEDICS CORPORATION MARINER PEDICLE SCREW SYSTEM; SET SCREW Back to Search Results
Model Number 41-1010
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the x-rays provided confirms levels l2-l5 were fused using seaspine's regatta interbody.It was also confirmed that the set screws in the iliac had separated from the construct.The first part of the revision occurred on (b)(6) 2022, which consisted of removing the seaspine reef to interbody from l5-s1 and replacing it with a seaspine vu apod prime nanometalene interbody to increase stability and decrease some of the load on the pedicle screws (and set screws).The second part of the revision occurred on (b)(6) 2022 to replace the mariner set screws that reportedly loosened for the second time.The mariner lateral connectors were also replaced with closed lateral connectors as additional supplementation to the existing construct.Review of labeling: possible adverse events: bending, disassembly or fracture of implant and components loosening of spinal fixation implants may occur due to, latent infection, and/or premature loading, possibly resulting in bone erosion, migration or pain.
 
Event or Problem Description
A patient underwent spinal surgery consisting of seaspine's mariner pedicle screw system and seaspine's regatta and reef to intervertebral fusion devices on (b)(6) 2020.On (b)(6) 2022, it was reported that a revision surgery occurred on (b)(6) 2020 to replace three mariner set screws that loosened in the postoperative period (ref: 3012120772-2022-00002 for this event).Two of the three set screws reportedly loosened again, prompting a second revision surgery, which occurred on (b)(6) 2022, prompting this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARINER PEDICLE SCREW SYSTEM
Common Device Name
SET SCREW
Manufacturer (Section D)
SEASPINE ORTHOPEDICS CORPORATION
5770 armada dr.
carlsbad CA 92008
Manufacturer (Section G)
SEASPINE ORTHOPEDICS CORPORATION
5770 armada dr.
carlsbad CA 92008
Manufacturer Contact
audrey mudderman
5770 armada dr.
carlsbad, CA 92008
7602165137
MDR Report Key13444889
Report Number3012120772-2022-00001
Device Sequence Number4492576
Product Code NKB
UDI-Device Identifier10889981155100
UDI-Public10889981155100
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K173882
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number41-1010
Device Catalogue Number41-1010
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Initial Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEASPINE'S REEF TO INTERBODY; SEASPINE'S REGATTA LATERAL INTERBODY; SEASPINE'S VU APOD PRIME NANOMETALENE INTERBODY
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-