| Model Number |
41-1010 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/28/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient underwent spinal surgery consisting of seaspine's mariner pedicle screw system and seaspine's regatta and reef to intervertebral fusion devices on (b)(6) 2020.On (b)(6) 2022, it was reported that a revision surgery occurred on (b)(6) 2020 to replace three mariner set screws that loosened in the postoperative period.Two of the three set screws reportedly loosened a second time, prompting a second revision surgery, which occurred on (b)(6) 2022, prompting report 3012120772-2022-00001.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the x-rays provided confirmed that the set screws in the iliac had separated from the construct.A revision surgery occurred on (b)(6) 2020 to reinstrument the failed hardware.Review of labeling: possible adverse events: bending, disassembly or fracture of implant and components loosening of spinal fixation implants may occur due to, latent infection, and/or premature loading, possibly resulting in bone erosion, migration or pain.
|
| |
|