Investigations: visual inspection: no significant deformations or damage of the valve, but during the inspection visible deposits were detected on the deflector and in the control reservoir.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: the results show that the valves operate within the accepted tolerance in both positions.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Klick force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, no deposits were found in both valves.Results: based on our investigation, we are unable to substantiate the clinical claim of under-drainage.At the time of our investigation, the valves were operating within the specified tolerances.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from (b)(4) kg.No further regulatory actions are required from our point of view.
|