• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; LIGATION / VESSEL CLIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL572T
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap ag that a ligature clip 12 mag.= 144 pcs (part # pl572t) was used during a procedure performed on (b)(6) 2022.According to the complainant, during the procedure, 5 millimeter (mm) clips were used with the device.The operator was able to place the first clip into the patient without issue, however, when the next clip was fire, the entire magazine came loose in the patient's abdomen.The magazine was retrieved, and reattached, but when it was re-tested outside the patient, the magazine came loose.A new clip magazine was then used.After attaching to the pliers and being careful to firmly seat the magazine, the device was tested outside the patient, but the issue recurred.The whole magazine pushed away from the pliers when the clips were fed and detached from the forceps.A third magazine was then able to be used to successfully complete the procedure.The complaint device was returned to the manufacturer for evaluation.An additional medical intervention was necessary.Although requested, additional information has not been made available.The adverse event is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap ag that a ligature clip 12 mag.= 144 pcs (part # pl572t) was used during a procedure performed on (b)(6) 2022.According to the complainant, during the procedure, 5 millimeter (mm) clips were used with the device.The operator was able to place the first clip into the patient without issue, however, when the next clip was fire, the entire magazine came loose in the patient's abdomen.The magazine was retrieved, and reattached, but when it was re-tested outside the patient, the magazine came loose.A new clip magazine was then used.After attaching to the pliers and being careful to firmly seat the magazine, the device was tested outside the patient, but the issue recurred.The whole magazine pushed away from the pliers when the clips were fed and detached from the forceps.A third magazine was then able to be used to successfully complete the procedure.The complaint device was returned to the manufacturer for evaluation.An additional medical intervention was necessary.Although requested, additional information has not been made available.The adverse event is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Investigation: the investigation was carried out visually and microscopically.Here we found broken off noses, broken off latches and clip jams.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(2) with this error pattern.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence 3(5)) according to din en iso 14971 is still acceptable.Explanation and rationale: according to the quality standard and dhr files a material defect and a production error can be excluded.There are no hints of a pre-damage or similar.If the cartridge is engaged not completely, has not been mounted correctly or was damaged during inserting, there is an impairment of product functionality.A too fast application, damaged cartridge during inserting or a cartridge which is engaged not completely could also lead to the described errors.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.According to the investigation results, a product safety case (psc) was initiated.Any action regarding capa will be addressed with this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGATURE CLIP 12 MAG.= 144 PCS.
Common Device Name
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key13445855
Report Number9610612-2022-00009
Device Sequence Number19956511
Product Code FZP
UDI-Device Identifier04038653341856
UDI-Public4038653341856
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K081031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2026
Device Model NumberPL572T
Device Catalogue NumberPL572T
Device Lot Number52699410
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer04/04/2022
Initial Report FDA Received Date02/03/2022
Supplement Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-