MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL FMS VUE PUMP-SHAVER BOX; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL FMS VUE PUMP-SHAVER BOX; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Model Number 284004
Device Problem Increase in Pressure (1491)
Patient Problem Extravasation (1842)
Event Date 01/19/2022
Event Type  Injury  
Event Description
It was reported by the affiliate in (b)(6) that during an unknown surgery on (b)(6) 2022, it was observed that the patient had a swollen knee after using fms vue pump-shaver box device that had a high pressure.The pressure was lowered to complete the procedure without delay.The current status of the patient was unknown.No additional information was provided.
 
Manufacturer Narrative
Udi: (b)(4).The lot was unknown.The lot number was unknown; therefore, the expiration date, manufacturing site name, and device manufacture date were unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d4: the serial number was reported as unknown on the initial report; and has been updated accordingly.Therefore, udi: (b)(4).
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H4: the device manufacture date was reported as unknown on the initial report; and has been updated accordingly.Investigation summary: the complaint device was not returned after multiple attempts for device return, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (b)(6) and no non-conformance's were identified.Should the device ever be received back in the future, this complaint file will be reopened at that time and an evaluation will be performed and documented.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent to which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP-SHAVER BOX
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle CH-24-00
SZ   CH-2400
6013142063
MDR Report Key13446095
MDR Text Key289269135
Report Number1221934-2022-00411
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705027934
UDI-Public10886705027934
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K171237
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number284004
Device Catalogue Number284004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-