• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BERLIN HEART GMBH BLOOD PUMP PU VALVES 30 ML, IN/OUT Ø9 MM; VENTRICULAR ASSIST DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BERLIN HEART GMBH BLOOD PUMP PU VALVES 30 ML, IN/OUT Ø9 MM; VENTRICULAR ASSIST DEVICE Back to Search Results
Model Number P30P-001X01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemolysis (1886)
Date of Event 01/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The site contacted berlin heart, inc on (b)(6) 2022 to report that a patient supported in the lvad configuration had developed hemolysis and was planning a pump change.The ldh levels were as follows: baseline 853 ((b)(6) 2021); 1362 ((b)(6) 2021); 1263 ((b)(6) 2021); 1253 ((b)(6) 2021); 2056 ((b)(6) 2021); 1429 ((b)(6) 2021); 2365 ((b)(6) 2022), 4183 ((b)(6) 2022).The high end of ldh normal for the site is 334 for this patient's age.Of note: this patient's baseline has always been elevated due to liver failure secondary to a failing fontan physiology.The site also reported that there was a soft "chirp" that was audible at the outflow valve since (b)(6) 2021 and that goes away with slight pressure over the pu valve.However, the sound returns when the pressure is released.This chirp has been unchanged since (b)(6) 2021.The pump has been filling and ejecting completely, and no dimples or graphite dust have been noted.
 
Additional Manufacturer Narrative
The excor blood pump, s/n, (b)(4) on the patient at the time of the event was in use from (b)(6) 2021 until (b)(6) 2022 (103 days).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD PUMP PU VALVES 30 ML, IN/OUT Ø9 MM
Common Device Name
VENTRICULAR ASSIST DEVICE
Manufacturer (Section D)
BERLIN HEART GMBH
wiesenweg 10
berlin, deu 12247
GM  12247
MDR Report Key13446386
Report Number3008454189-2022-00003
Device Sequence Number4328138
Product Code DSQ
UDI-Device Identifier04260090040126
UDI-Public04260090040126
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/03/2022
Report Date (Section F) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model NumberP30P-001X01
Device Catalogue NumberP30P-001X01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/18/2022
Device Age1 YR
Event Location Hospital
Date Report to Manufacturer01/18/2022
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/03/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age7 YR
Patient SexMale
Patient Weight21 KG
Date Report Sent to FDA02/03/2022
-
-