AVANOS MEDICAL INC. ON-Q PAIN RELIEF SYSTEM WITH SOAKER CATHETER, 100 ML, 2 ML/HR, 5 IN./12.5 CM; KITS
|
Back to Search Results |
|
| Model Number |
PS12502 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 03-feb-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Fill volume: 115 ml, flow rate: unknown , procedure: unknown, cathplace: unknown, date/time when infusion started: (b)(6) 2021, date/time when infusion ended: (b)(6) 2021.It was reported that all the medicine infused in 18-hours, earlier than expected.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for the reported lot number 20043936, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The pump was received partially full, it was drained and refilled for evaluation.The evaluation noted the device was within specifications, the reported incident was not replicated during the sample evaluation.The investigation determined that the returned device functioned as intended during the sample evaluation, no defect was found; therefore, complaint is not confirmed.The investigation evaluation noted the flow rate of a pump could be affected by different external factors such as the increase of temperature and the pump position, the user and facility conditions were unknown, therefore it is not possible to know if that could have contributed to the failure.Per the instructions for use (ifu) the flow rate may be higher or lower at the first and last hours of infusion, and the labeled flow rates are based on the use of normal saline as diluent, if a different diluent/drug is used the viscosity could be affected and cause the flow rate to increase.Root cause could not be determined.All information reasonably known as of 16-mar-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|