| Catalog Number |
51-100080 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed a follow-up mdr will be submitted report source: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported by the distributorship that peeling was discovered in the sealing area of the products.No adverse events were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual inspection of the returned product found a void in the tyvek seal that is greater than 6mm acceptance criteria.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|