• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ASSISTANT WITH STABLESOFT 3S; UNIVERSAL STABILIZER ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ASSISTANT WITH STABLESOFT 3S; UNIVERSAL STABILIZER ARM Back to Search Results
Model Number T401210S
Medical Device Problem Codes Break (1069); Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.
 
Event or Problem Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, part of the retractor came off.Per facility, a plastic dot came off the paddle onto to the heart and was seen by the first assistant noted a small white foreign body in the surgical field on the heart.The first assistant was able to visualize and retrieve the small object as they were wearing a pair of magnifying surgical loupes.Per clinician, the foreign object was identified as possibly being a small piece of silicone over molding from the retractor.The product was not changed out.The procedure was completed successfully.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 4238, 4315).Type of investigation: #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation finding #: 4238 - contamination of environment by device.Investigation conclusions: 4315 - cause not established.Upon evaluation of the returned device, the foreign material is confirmed; however, it was unable to be confirmed where on the device the foreign material derived from.The supplier was notified.A retention sample from the same product code and lot number was obtained and visually inspected.No foreign materials were found anywhere on the device.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on february 3, 2022.Upon further investigation of the reported event, the following information is new and/or changed: g6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSISTANT WITH STABLESOFT 3S
Common Device Name
UNIVERSAL STABILIZER ARM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key13446467
Report Number1124841-2022-00019
Device Sequence Number4321693
Product Code MWS
UDI-Device Identifier00699753450516
UDI-Public(01)00699753450516
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberT401210S
Device Catalogue NumberN/A
Device Lot NumberZG07A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer02/02/2022
03/31/2022
Initial Report FDA Received Date02/03/2022
Supplement Report FDA Received Date02/16/2022
04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-