CARDINAL HEALTH CURITY MONO-FLO 14FR WO PVPI; COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER Back to Search Results
Model Number 6162MF
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that after the balloon was inflated for testing, it did not deflate.
 
Additional Manufacturer Narrative
H6 investigation findings: 4247 - root cause undetermined.Investigation summary: the device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported issue.The dhr showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.One decontaminated device was received for evaluation.A visual and functional inspection was carried out per our procedure.The reported condition was observed; the catheter balloon presented difficulty when attempted to remove water with a syringe.The affected component is produced by an external supplier; our assembly process consists solely of a collection and placement operation of this material therefore the exact root cause was unable to be determined.The supplier was notified and awareness training for this type of issue will be conducted.There is no further action required at this time.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURITY MONO-FLO 14FR WO PVPI
Common Device Name
COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER
Manufacturer (Section D)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill sariava
777 west street
mansfield, MA 02048
5086183640
MDR Report Key13446500
Report Number9612030-2022-03157
Device Sequence Number19913729
Product Code KNX
UDI-Device Identifier10884521050587
UDI-Public10884521050587
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number6162MF
Device Catalogue Number6162MF
Device Lot Number2104826564
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer02/01/2022
Initial Report FDA Received Date02/03/2022
Supplement Report FDA Received Date05/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-