The device is returned and an evaluation completed for it.In addition the customer returned an unused device from the same lot.Upon inspection and testing of the used device, the cutting wire was found to be broken off at the proximal end.The broken wire sticking out at the distal end measured about 23 mm long, and its tip was charred and melted indicating thermal damage.The coated portion covering the knife was torn at tip but present.The remaining wire retracted inside the working tube sheath.There is no abnormality on the working section, as its handle is functioning properly.The remaining wire inside the tube sheath moves smoothly back and forth when the handle is operated.In addition, the returned new device was inspected and found no abnormalities on the appearance.The cutting wire is in perfect condition.The device was checked with test generator esg-400, setting on monopolar cut mode at 120 w.Observed that the cutting wire could generate energy output and cut a piece of tissue sample without a problem.The cutting wire stayed intact and did not snap after multiple activation attempts.The new device passed functional inspection.The reported complaint was not duplicated on the new device.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Correction to g3 of the initial medwatch.The aware date should be 24-nov-2021.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the confirmation result and the investigation results in the past, a likely mechanism causing the broken cutting wire might be the following: 1.The cutting wire at a torn area of the coated portion came into contact with the distal end of the endoscope while the forceps elevator was raised.2.The output was activated in state of ¿1¿ description, and the cutting wire became hot instantly.This caused the cutting wire to break.It has been confirmed that the tear of the coated portion of the cutting wire could duplicate by the following mechanism: 1.The forceps elevator of the endoscope was raised.2.When the cutting wire deflects, the coated portion of the cutting wire and the metal part of the distal end of the endoscope come into contact.3.In state of description ¿2¿, the cutting wire was moved back and forth.This caused the coated portion of the cutting wire to tear.The slider was pushed more than needed.This caused the cutting wire to deflect.The event can be prevented by following the instructions for use which state: since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.Olympus will continue to monitor field performance for this device.Investigation activities have been opened to manage the actions related to this report and any required mdr reporting.
|