| Catalog Number |
2C8720 |
| Medical Device Problem Codes |
Backflow (1064); Filling Problem (1233)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).Device manufacturer address 1: (b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was backflow of transfusion blood while using a clearlink system y-type blood/solution set.This was further described as "while infusing blood the roller clamps on the y-tube failed allowing the backflow of the blood into the 0.9% ns (normal saline) bag as well as allowing 0.9% ns to leak into the blood administration increasing the total volume of infused fluid." this issue was identified during a patient blood infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the actual sample was received for evaluation.Visual inspection was performed using the naked eye which did not identify any abnormalities that could have contributed to the reported conditions.All components were correctly placed and according to specifications.Functional testing was performed including priming test, clear passage and pressure testing; and no issues were observed; however, a cut was observed on the tubing.The reported backflow issue could not be verified.The cause of the cut was due to a manufacturing process related issue; a potential root cause was determined as an incorrect tubing placement as part of the coiling operation, leading the packaging machine to cut the tubing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|