Model Number IPN043837
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).
 
Event or Problem Description
It was reported that "the material of several devices of this lot is breaking.Defect was noted prior to use".No patient involvement reported.
 
Additional Manufacturer Narrative
Qn#(b)(4).Twenty representative samples were returned by the customer for investigation.A visual exam was performed and no defects were observed.The manufacturing site reports that in the current manufacturing procedure 100% leak testing is conducted during the assembly process and a 100% visual inspection is also conducted at the packing area; thus any defective product will be detected prior to release.The manufacturing site also reports that ten pieces of this product code were taken from current production at the manufacturing facility.A visual exam was conducted along with leak testing.No issues were detected.All ten samples passed the leak test.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "the material of several devices of this lot is breaking.Defect was noted prior to use".No patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON HUMID-VENT FLEX TUBE
Common Device Name
SET, TUBING AND SUPPORT, VENTI
Manufacturer (Section D)
TELEFLEX MEDICAL SDN. BHD.
perak, west malaysia
Manufacturer (Section G)
TELEFLEX MEDICAL SDN. BHD.
lot no : pt2577 jalen perusahaan
4 kamunting industrial estate
perak, west malaysia 34600
MY   34600
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key13446668
Report Number8040412-2022-00023
Device Sequence Number4321724
Product Code BZO
UDI-Device Identifier04026704388769
UDI-Public04026704388769
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date04/30/2026
Device Model NumberIPN043837
Device Catalogue Number20011
Device Lot NumberKMH21F0394
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2022
Supplement Date Received by Manufacturer03/04/2022
Initial Report FDA Received Date02/03/2022
Supplement Report FDA Received Date03/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-