• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2021
Event Type  malfunction  
Event Description
The sheath subsequently tested out of specification per the manufacturer's investigation.
 
Manufacturer Narrative
Product event summary: the 4fc12 sheath with lot 0010805965 was returned and analyzed.Visual inspection was performed on the shaft identified a shaft kink/twist at approximately 2.44 inches from the tip.Functional testing of the steering mechanism and a deflection test were performed.The shaft would bend as initially intended and would bend in the plane.There was no difficulty, no friction, or any noise in the steering mechanism.In conclusion, the sheath failed the returned product inspection due to a kink/twist that was observed on the shaft.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13446987
MDR Text Key288696573
Report Number3002648230-2022-00051
Device Sequence Number1
Product Code DRA
UDI-Device Identifier00763000418052
UDI-Public00763000418052
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/24/2023
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010805965
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2021
Date Manufacturer Received01/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-