Model Number Z9002
Medical Device Problem Codes Break (1069); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during insertion the intraocular lens (iol) haptic broke and jammed in the pscst30 cartridge; account stated possible use error and folding issue.Another lens with the same model and diopter was implanted in the patient¿s ocular dexter (right eye).There was no medical and no surgical intervention.No further information is available.
 
Additional Manufacturer Narrative
Patient weight, ethnicity and race: unknown as information was asked but not provided.Date implanted: not applicable as the iol was not implanted.Date explanted: not applicable as the iol was not implanted; therefore, not explanted.Concomitant medical products: pscst30 cartridge lot unknown, ovd, balanced salt solution (bss) & si40.Device evaluated by mfr: the device was not returned for analysis therefore, a visual analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts were made to contact the customer account requesting additional information regarding complaint however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Product investigation was completed, and the following fields were updated accordingly: section d-9: device available for evaluation: yes section d-9: date returned to manufacturer: 25-mar-2022 section h-3: device evaluated by manufacturer: yes device evaluation: visual inspection under magnification revealed that the lens was received with a lens detached.The lens was cleaned, and lens damage and scratches could be observed on the lens.No further issues could be identified with the lens.The complaint issues (haptic damaged and stuck in cartridge) were not confirmed.However, the observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.The reported issues were not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Corrected data: section h6: as part of an internal review of our mdrs it was identified that the code "medical device problem code: 42983 - mechanical jam" was in advertently not entered in the section "h6" of the initial report.Therefore, the correction has been included in this supplemental mdr report as indicated below: section h6: medical device problem code: 42983.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13447676
Report Number3012236936-2022-00288
Device Sequence Number14793153
Product Code HQL
UDI-Device Identifier05050474530706
UDI-Public(01)05050474530706(17)240920
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2024
Device Model NumberZ9002
Device Catalogue NumberZ900200205
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer05/18/2022
10/19/2022
Initial Report FDA Received Date02/03/2022
Supplement Report FDA Received Date05/25/2022
10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-