MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105ELGYNA ELI LILY GRAY; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105ELGYNA ELI LILY GRAY; SYRINGE, PISTON Back to Search Results
Model Number MMT-105ELGYNA
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Insufficient Information (4580)
Event Date 11/26/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Serial number: n/a; software version: n/a ; color: grey; battery life remaining: n/a.Customer reports: the dial is very hard to push past 4 units, if he turns really hard it will go past 4 but if he tries to dispense insulin nothing will come out.Per visual inspection: missing dose window.Several attempts were made to pair inpen, every time app displayed dose doesn't match and inpen not found.The inpen does not pair with commercial mobile app.The inpen screw retracts when dialing and advances when turning dose knob and high resistance while dispensing and dialing noted.The dose button was removed and no dust / debris under the dose button or dose knob noted.Inpen was cut open and after inspection it was found that the encoder pattern wheel tabs rotating and traveling off the keyed slots of dose nut guides.Encoder pattern wheel should never rotate.This causes an unexpected travel of the encoder pattern wheel creating resistance to dial and dispensing.Also contamination build up found caused by encoder wheel tabs rubbing at the walls of the dose nut.A battery test was performed, and battery measured 3.0 volts.In conclusion: it was determined that pairing anomaly and the customer complaint of hard to turn the inpen, hard to dial, hard to push to give insulin was confirmed due to encoder pattern wheel rotating and traveling against the walls of the dose nut.
 
Event Description
Information received by medtronic indicated that the in pen had hard to push past 4 unit and customer stated they watched lead screw.No harm requiring medical intervention was reported.The insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105ELGYNA ELI LILY GRAY
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
tricha miles
1223 world trade drive, 100
san diego, CA 92128
7635140379
MDR Report Key13448528
MDR Text Key287531069
Report Number3012822846-2022-00083
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000320
UDI-Public(01)000010862088000320(17)230419
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/19/2023
Device Model NumberMMT-105ELGYNA
Device Catalogue NumberMMT-105ELGYNA
Device Lot NumberB0061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2021
Date Manufacturer Received01/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age52 YR
-
-