Model Number AU00T0
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device with the lens was returned loose inside the opened carton.The plunger lock and lens stop were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was at the start point of the loading area.The lens was located in the loading area with no signs of advancements.There was no evidence observed of a plunger override or a retraction of the plunger.There was no ovd travel path observed.The anterior of the nozzle tip was bent down, possibly due to being returned loose in the carton.Product history records were reviewed and documentation indicated the product met release criteria.Viscoelastic was not provided.It is unknown if the qualified product was used.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non healthcare professional reported during an intraocular lens implant procedure, the lens did not eject.Patient contact was noted but no patient harm was observed.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13448941
Report Number1119421-2022-00175
Device Sequence Number5461810
Product Code KYB
UDI-Device Identifier00380652394499
UDI-Public00380652394499
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/05/2023
Device Model NumberAU00T0
Device Lot Number15105317
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/13/2022
Initial Report FDA Received Date02/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-