| Model Number |
200-02-35 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/09/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 10.09 years since initial implant, this male patient with bmi 22.7 underwent a left side tka and was implanted with an exactech uhmwpe insert, serial # (b)(4).On (b)(6) 2021, the patient underwent revision due to aseptic loosening.No additional information is available or forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the revision reported was likely the result of an insufficient bond between the implant and the bone, which led to aseptic (non-infected) loosening.However, this cannot be confirmed as the devices were not available for evaluation and pre-revision x-rays were not provided.(d10) concomitant device(s): 200-02-35, 2035253 - three peg patella 35mm section h11: *the following sections have corrected information: (d4) catalog number: 02-012-35-4015, serial number: (b)(6), expiration date: 20-may-2017, unique identifier (udi) #: (b)(4).(d10) concomitant medical products: see h10.(g4) pma/510(k)number: k033883.(h4) device manufacture date: 29-oct-2009.
|
| |
|