| Model Number |
208-25-18 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/17/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 9.68 years since initial implant, this male patient with bmi 32.1 underwent a right side tka and was implanted with an exactech uhmwpe insert, serial # (b)(4).On (b)(6) 2019, the patient underwent revision due to aseptic loosening.No additional information is available or forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the revision reported was likely the result of an insufficient bond between the implant and the bone and patient-related conditions, which led to aseptic (non-infected) loosening.However, this cannot be confirmed as the devices were not available for evaluation and pre-revision x-rays were not provided.
|
| |
|