WRIGHTS LANE SYNTHES USA PRODUCTS LLC ROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS; PLIERS, SURGICAL Back to Search Results
Model Number 388.50
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2021, the patient underwent a spinal fusion procedure.Two uss persuaders were unable to be used for the procedure due to mechanical issues.It is unknown how the procedure was completed.Patient status is unknown.This report is for one (1) rod introduction pliers for side-opening implants this is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Occupation: initial reporter is synthes sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d4 d10: date of concomitant therapy is (b)(6) 2021.H3, h4, h6: part: 388.50 lot: t173508 manufacturing site: tuttlingen release to warehouse date: may 10, 2019 a review of the device history records was performed for the finished device lot number, and no non-conformances were identified.Service and repair evaluation: the repair technician reported the device has dents on the nut and the tube and collet don¿t slide smoothly.Worn out parts is the reason for repair.The cause of the issue is unknown.The following parts were replaced: spring, tube, nut, collet and retaining collar.The item was repaired per the inspection sheet, passed synthes final inspection and will be returned to the customer upon completion of the service and repair process.Finalized service record will be archived.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS
Common Device Name
PLIERS, SURGICAL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen 78532
GM   78532
Manufacturer Contact
dan nelson
375 river park circle
marquette, MI 49855
MDR Report Key13450544
Report Number2939274-2022-00355
Device Sequence Number4675778
Product Code HTC
UDI-Device Identifier10705034773292
UDI-Public(01)10705034773292
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number388.50
Device Catalogue Number388.50
Device Lot NumberT173508
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer02/07/2022
03/17/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date03/07/2022
04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/10/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ROD INTRODUCTN PLIERS F/ SIDE-OPEN IMPS.
Patient SexUnknown
-
-